DINAMAP ProCare Monitor Service Manual
This service manual covers the DINAMAP ProCare Patient Monitor, specifically models 100, 200, 300, and 400, with or without integrated printers. The document provides detailed instructions for service technicians regarding the configuration, periodic maintenance, and safety testing of the device. It includes specific procedures for parameter tests such as blood pressure, pulse, temperature, and SpO2, as well as leakage tests and communication port verification.
| Operating Temperature | +5° C to + 40° C (-41 F to +104 F) |
|---|---|
| Storage Temperature | -20° C to +50° C (-4 F to +122 F) |
| Power Requirements (AC) | 120 VAC/60 Hz |
| Power Requirements (DC) | 24 VDC (nominal), at 1A |
| Battery Capacity | 6 volt, 3.3 amp-hours |
| Minimum Operation Time | 2 hrs |
| Dimensions (H x W x D) | 9.8 in x 9.8 in x 6.9 in (25.0 cm x 24.8 cm x 17.5 cm) |
| Weight | 7.8 lb (3.5 Kg) |
| NIBP Cuff Pressure Range (Adult) | 0 mmHg to 290 mmHg |
| SpO2 Range | 70-100% |
This manual applies to ProCare Monitor models 100, 200, 300, and 400, with or without printers.
The minimum operation time is 2 hours under specific conditions including a 5-minute cycle with an adult cuff.
When external DC power becomes available, the system rapidly switches from battery power to external power.
The NIBP accuracy meets the AAMI/ANSI standard SP-10 with a mean error of ±5 mmHg.