GE Medical Systems ProCare Patient Monitor Service Manual

DINAMAP ProCare Monitor Service Manual


GE Medical Systems ProCare Patient Monitor Service Manual - cover page
Brand
GE Medical Systems
Category
Other products
Document type
Service Manual
Language
English
Pages
95
File format
PDF
File size
4.8 MB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
95B1CA418EA1EC7CBCA10931649C4B96

About this manual

This service manual covers the DINAMAP ProCare Patient Monitor, specifically models 100, 200, 300, and 400, with or without integrated printers. The document provides detailed instructions for service technicians regarding the configuration, periodic maintenance, and safety testing of the device. It includes specific procedures for parameter tests such as blood pressure, pulse, temperature, and SpO2, as well as leakage tests and communication port verification.

What's inside

  • Introduction
  • Product Description
  • Principles of Operation
  • Calibration & Maintenance
  • Safety Testing
  • ProCare Patient Monitor Parameter Tests
  • Alarm Code Interpretation
  • Assembly Drawings & Schematics

Specifications

Operating Temperature+5° C to + 40° C (-41 F to +104 F)
Storage Temperature-20° C to +50° C (-4 F to +122 F)
Power Requirements (AC)120 VAC/60 Hz
Power Requirements (DC)24 VDC (nominal), at 1A
Battery Capacity6 volt, 3.3 amp-hours
Minimum Operation Time2 hrs
Dimensions (H x W x D)9.8 in x 9.8 in x 6.9 in (25.0 cm x 24.8 cm x 17.5 cm)
Weight7.8 lb (3.5 Kg)
NIBP Cuff Pressure Range (Adult)0 mmHg to 290 mmHg
SpO2 Range70-100%

Frequently asked questions

Which ProCare monitor models does this manual cover?

This manual applies to ProCare Monitor models 100, 200, 300, and 400, with or without printers.

What is the minimum operation time of the battery?

The minimum operation time is 2 hours under specific conditions including a 5-minute cycle with an adult cuff.

How does the monitor handle power switching?

When external DC power becomes available, the system rapidly switches from battery power to external power.

What is the accuracy of the NIBP measurement?

The NIBP accuracy meets the AAMI/ANSI standard SP-10 with a mean error of ±5 mmHg.