This document is the T/Pump service manual, which introduces the features, functions, specifications, usage methods of T/Pump
This service manual provides instructions for qualified medical service personnel to repair and perform function tests on the T/PUMP TP600 and TP700 series localized temperature therapy systems. It includes safety precautions, system introductions, and detailed features of the control units and pads.
| Reservoir capacity | 93 oz (2.75 l) maximum |
|---|---|
| Flow rate | 9 gph (34 lph) minimum with pad attached |
| Ambient operating temperature | 60°F to 90°F (15.6°C to 32.2°C) |
| Maximum Contact Surface Temperature | 107°F (42°C) |
| High Limit Safety Temperature | 110°F to 120°F (43.3°C to 49°C) |
| Average temperature accuracy | ±2°F at 107°F (±1°C at 42°C) |
Only qualified medical service personnel should repair or perform function tests on the T/Pump.
The setpoints are 107°F (42°C) for high heat, 100°F (38°C) for low heat, and 50°F (10°C) for cooling.
Press and hold the setpoint button for 2 seconds to lock or unlock the setpoint.