This file is the service information sheet for the T5-600 and T5-1000 Electrical Testers. It introduces the following information for these products: Safety information, Parts and service information, Specifications, Cleaning procedure, Required equipment, Performance tests, Parts and accessories list, Battery replacement procedure
This service manual provides safety guidelines, specifications, and procedures for servicing the Fluke T5-600 and T5-1000 electrical testers. It includes safety warnings, calibration details, voltage and current ratings, cleaning instructions, and step-by-step performance tests for verifying accuracy. The manual also lists required equipment and outlines procedures for testing AC current, voltage functions, and hazardous voltage indicators.
The document is intended for qualified technicians and includes technical specifications for both models, including accuracy, temperature ranges, battery life, and safety certifications. It covers both serial number ranges and provides detailed test procedures with calibration tables and component requirements.
| Maximum Voltage (Terminal to Ground) | T5-600: 600 V rms, Overvoltage Category III; T5-1000: 1000 V rms, Overvoltage Category IV |
|---|---|
| Battery Type | AA (2) |
| Accuracy (Voltage) | 1V +(1.5% + 2 digits) for T5-600; 1V +(1% +1 digit) for T5-1000 |
| Operating Temperature | -10 °C to +50 °C (14 °F to 122 °F) |
| Surge Protection | T5-600: 6 kV per IEC 1010-1, 1990-09; T5-1000: 8 kV per IEC 1010-1, 1990-09 |
The T5-1000 has a maximum voltage rating of 1000 V rms, Overvoltage Category IV.
The T5-600 and T5-1000 use AA batteries; alkaline batteries last 360 hours continuously, and zinc chloride batteries last 125 hours.
The accuracy for voltage measurements on the T5-600 is 1V +(1.5% + 2 digits).
The testers comply with ANSI/ISA-S82.01-94, UL, CSA/CAN C22.2 No.1010.1-92, and EN61010-1 1993.