AD 131 and AD 132 pH/mV/Temperature Waterproof Meters USER MANUAL
The AD 131 and AD 132 are portable waterproof meters designed for pH, ORP, and temperature measurements in harsh industrial environments. They feature 7 memorized buffer points for calibration, automatic or manual temperature compensation, and GLP compliance. The AD132 includes a PC interface and printer support. Both models come with pH electrodes, temperature probes, batteries, and a user manual. Calibration, operation, and troubleshooting are covered in detail, with warranty information for 2 years on the meter and 6 months on probes.
| Range | -2.00 to 16.00 pH ±2000 mV -20.0 to 120.0 °C |
|---|---|
| Resolution | 0.01 pH 0.1 mV (±999.9 mV) / 1 mV (outside) 0.1 °C (0.1 °F) |
| Accuracy | ±0.01 pH (@25 °C/77 °F) ±0.2 mV up to ±699.9 mV ±0.5 mV up to ±999.9 mV ±2 mV (outside) ±0.4 °C (±0.7 °F) excluding probe error |
| pH Calibration | Up to 5-point calibration, with 7 standard buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) |
| Temperature Compensation | Manual or automatic, -20.0 to 120.0 °C (-4.0 to 248.0 °F) |
| Battery Life | Approx. 300 hours of continuous use |
| Dimensions | 188 x 96 x 70 mm |
| Weight | 460 g |
The meter is warranted for a period of 2 years, while probes are warranted for 6 months.
They include 7 memorized buffers for pH calibration, LCD messages for easy calibration, manual or automatic temperature compensation, calibration time-out alarm, log-on-demand, HOLD feature, GLP feature, and PC interface (AD132 only).
They come with an AD1230B pH electrode, AD7662 temperature probe, 4 x 1.5V AA alkaline batteries, user manual, and (AD132 only) a connection for external serial printer.
Use up to 5-point calibration with 7 standard buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45).