Precision Radiation Instruments Inc MODEL 121 Instruction Manual

Model 121 is a Geiger counter with three probes: Type A probe contains a thin mica end-window Geiger tube, used for detection of alpha and beta particles; Type B probe is used for detection of beta and gamma radiation in general prospecting, and has a sliding shield over the side window for exclusion of beta particles; Type D probe is a sealed probe, with a long cable, for examination of drill-holes up to 1,000 feet in depth.


Precision Radiation Instruments Inc MODEL 121 Instruction Manual - cover page
Brand
Precision Radiation Instruments, Inc.
Document type
User Manual
Language
English
Pages
9
File format
PDF
File size
2.1 MB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
6FF9DEEA41829105B5BCF786F86E5CA8

About this manual

This manual provides operating instructions for the Model 121 Geiger Counter, including calibration procedures for different probe types and time constant settings. It also details how to connect external equipment like recorders and audio monitors, and lists authorized service stations for repairs. The document includes information on warranty and return procedures, emphasizing that service must be performed by authorized stations to maintain warranty validity.

What's inside

  • Calibration
  • Probe Types
  • Time Constant Settings
  • Recorder Connection
  • Audio Monitor Connection
  • Guarantee and Service
  • Authorized Service Stations

Frequently asked questions

How do you calibrate the Type "D" Drill Hole Probe?

Hold the calibrated radium disk flat against the probe about 1-3 inches from the tip, move it slowly to find the spot giving maximum reading at the 2.0 MR per HR position, and adjust the "CALIB." control until the average reading matches the disk’s stamped value.

What is the recommended time constant setting for drill hole work?

Use the "FAST" time constant setting for drill hole work to detect large changes in radioactivity and to allow faster probe movement.