About this manual

The MACS epic CPAP System is a lightweight, pneumatic device designed for delivering continuous positive airway pressure to spontaneously breathing patients. It operates using oxygen or compressed air and can deliver oxygen at adjustable levels from 21% to 100%. The system is intended for use by trained medical professionals in hospital and pre-hospital settings, including transport and emergency care. It includes safety features like alarms for low gas supply and disconnect events, and requires specific patient circuits for safe operation.

The manual provides detailed instructions for setup, patient connection, operational verification, troubleshooting, and maintenance. It includes warnings about incompatible gases, latex-free components, and the need for authorized servicing. The device is not designed for use with mechanical ventilation or flammable anesthetics.

What's inside

  • General Description
  • Warnings, Cautions, Notes
  • Controls and Patient Safety Systems
  • Operating Instructions
  • Patient Circuit
  • Accessories
  • Theory of Operation
  • Troubleshooting
  • Cleaning and Maintenance
  • Specifications

Specifications

Oxygen Delivery Range21% to 100%
Oxygen Only Operation65% oxygen
Driving Gas Requirements55 + 15 psi (380 + 100 kPa), 40 liter/minute minimum
Maximum Altitude15,000 feet

Frequently asked questions

What is the oxygen delivery range of the MACS epic CPAP System?

The delivered oxygen is adjustable from 21 to 100 percent.

Can the MACS epic CPAP System be used with oxygen only?

Yes, it can be operated from oxygen only source gas with delivered oxygen at 65 percent.

What are the gas pressure requirements for the MACS epic CPAP System?

Driving gas requires 55 + 15 psi (380 + 100 kPa), with a minimum flow of 40 liters/minute.

What is the maximum altitude at which the MACS epic CPAP System can operate?

The system operates normally up to 15,000 feet.