This is an operation manual for a reference measurement instrument
This manual provides step-by-step instructions for operating the ReferenceMate REF2500 handheld shaker, including mounting procedures, power settings, frequency selection, and testing triaxial transducers. It also includes safety warnings and details on battery use and environmental protection. The device is designed for field use to verify sensor performance and cabling integrity.
The REF2500 operates at three frequencies (61.4 Hz, 100.0 Hz, 159.2 Hz) and supports both RMS and peak measurement modes. It features LEDs to indicate battery status and overload conditions. The unit is not certified for use in hazardous environments and requires careful handling to avoid damage to its internal components.
| Maximum Test Weight | 250 grams |
|---|---|
| Battery Type | Four AA alkaline batteries (rechargeable and lithium batteries may also be used) |
| Auto Shutdown | 90 seconds after power-on if no buttons are pressed |
To change the measurement type, press and hold the FREQ button for approximately 5 seconds.
If the unit is overloaded, the OOR LED will illuminate, indicating the maximum load has been exceeded.
Yes, rechargeable and lithium batteries may be used, but never mix battery types and always replace all four cells when changing batteries.