IM7264B is a instruction manual for model 7264B accelerometer
This manual provides step-by-step instructions for installing and testing the IM Sensor 7264B piezoresistive accelerometer. It includes precautions for handling, required mounting materials, and procedures for impedance, insulation resistance, and zero measurand output tests. The document also covers mounting torque specifications, cable installation, and recalibration intervals.
Additional guidance is provided for cleaning the unit and troubleshooting, with contact information for Endevco’s Application Engineering for further assistance.
| Mounting Torque | 2.6 in-lbf (0.29 N-m) recommended, 3.0 in-lbf (0.34 N-m) maximum |
|---|---|
| Zero Measurand Output (ZMO) | Within data sheet specified limits |
| Insulation Resistance | 100 megohms minimum |
Materials include an Allen wrench, flat washers, and socket head cap screws, all supplied with the unit.
The recommended torque is 2.6 in-lbf (0.29 N-m), with a maximum of 3.0 in-lbf (0.34 N-m).
Recalibration should be performed every 6 to 12 months, depending on usage.