Sensortechnik und Elektronik Pockau GmbH SM 5 D Survey Meter User Manual

This is a hand-held radiation meter for measuring the local dose rate of gamma radiation, X-rays as well as the area activity of beta and alpha radiation.


Sensortechnik und Elektronik Pockau GmbH SM 5 D Survey Meter User Manual - cover page
Brand
Sensortechnik und Elektronik Pockau GmbH
Document type
User Manual
Language
English
Pages
13
File format
PDF
File size
616 KB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
1FDE1185900AE9AABF56EF336ECB7F1B

About this manual

The SM 5 D is a handheld, battery-powered radiometer for measuring gamma, x-ray, beta, and alpha radiation. It features a display range of 0.00 to 299 μSv/h for dose rate and can measure area activity. The device includes safety instructions, measurement modes, and decontamination procedures. It is designed for use in medicine, industry, and research environments.

The manual covers setup, operation, and maintenance, including battery replacement, averaging measurements, and handling contamination. It also includes technical specifications, measuring principles, and alarm messages for overload conditions.

What's inside

  • Safety Instructions
  • Design and Control Elements
  • Technical Data
  • Measuring Principle
  • Performing Measurements
  • Evaluation of Measurement Results
  • Alarm Messages
  • Shipment
  • Maintenance
  • Decontamination

Specifications

Display Range (Dose Rate)0.00 μSv/h … 299 μSv/h
Energy Range (Photons)10 keV … 2 MeV
Weight400 g
Dimensions152 mm x 83 mm x 35 mm
Operating Temperature0°C to +50°C
Power Supply2 x 1.5 V (AA batteries)

Frequently asked questions

What is the display range for dose rate measurements?

The display range is 0.00 μSv/h to 299 μSv/h.

How do I activate the background lighting?

Press the power switch to turn on the device; the background lighting turns on for 10 seconds and can be reactivated by pressing the switch again.

What happens if the device is overloaded?

The display shows an overload message and the buzzer activates with a continuous tone.

How do I decontaminate the device?

Clean the device with a damp cloth; do not use solvents like gasoline or acetone.