T3-1 TUF-2000 Series Ultrasonic Flow/Heat Meter User Manual

This user manual provides detailed installation and operation instructions for TUF-2000 Series Ultrasonic Flow/Heat Meters


T3-1 TUF-2000 Series Ultrasonic Flow/Heat Meter User Manual - cover page
Brand
T3-1
Document type
User Manual
Language
English
Pages
35
File format
PDF
File size
1.7 MB
Published
18 October, 2023
Updated
18 October, 2023
MD5 checksum
6D6943BF7944F1100C99C9C47E57295E

About this manual

This manual provides step-by-step instructions for installing and operating T3-1's TUF-2000 Series ultrasonic flow and heat meters. It covers different converter types (separated, fixed, module) and transducer types (clamp-on, insertion, inline), along with wiring diagrams and setup procedures. The guide also explains how to check signal strength and transmission time ratio for proper installation.

The manual includes sections on mounting, wiring, parameter setup, and troubleshooting, with specific instructions for each type of installation. It emphasizes using manufacturer-provided cables and keeping cable lengths under 200m for optimal performance.

What's inside

  • Products Categories
  • Check Components
  • Measuring Diagrams
  • Converter Installation and Wiring Diagram
  • Transducer Introduction and Wiring Diagram
  • Display and Operation
  • Transducers Installation
  • Finish Installation

Frequently asked questions

What is the TUF-2000 Series used for?

It measures velocity of relatively clean liquids in full pipes using transit-time principle.

What are the components of an Ultrasonic Flow meter?

It consists of a Converter and a transducer.

What is the acceptable range for transmission time ratio during installation?

It should be between 97% and 103%.

What is the minimum signal strength required for proper operation?

Signal strength must be greater than 60.0.

Download PDF (1.7 MB)