This manual is a user manual for Digital Synthesis, High Reliability, Xenon Tube STROBOSCOPE DT-2349
This is the user manual for the DT-2349 digital stroboscope, a high-reliability instrument with xenon flash for measuring rotational speed. It features a digital display, microprocessor control, and adjustable settings via knobs and buttons. The device operates on AC 110-230V or via included DC adapter, with optional battery power. Safety warnings cover flash tube replacement, electrical hazards, and viewing precautions.
The manual covers operation, power supply, measuring procedures, and flash tube replacement. It includes specifications such as 0.1 RPM resolution under 1,000 RPM, 1 RPM above, and a 5-digit LCD display. Environmental conditions and duty cycle limits are specified for safe use.
| Display | 5 digits (0 to 99999) LCD |
|---|---|
| Flash Range | 60 to 32,000 RPM/FPM |
| Resolution | 0.1 RPM (<1,000 RPM), 1 RPM (≥1,000 RPM) |
| Accuracy | + (0.05% + 1d) |
| Power Supply | 110 Vac ±10%, 50/60 Hz or 220 Vac ±10%, 50/60 Hz or 230 Vac ±10%, 50/60 Hz |
| Operating Temperature | 0 to 50 °C (32 to 122 °F) |
| Operating Humidity | Less than 80% R.H. |
| Dimensions | 21 x 12 x 12 cm (8.3 x 4.8 x 4.8 inch) |
| Weight | 1 kg (2.2 lb) |
| Flash Duration | Approximately 5 to 12 microseconds |
| Flash Energy | 4 Watts-seconds (joules) |
| Flash Color Temperature | Xenon white 6,500 K |
| Beam Angle | 80 degrees |
| Duty Cycle | <2000 RPM: 2 hours, 2001-3600 RPM: 1 hour, 3601-8000 RPM: 30 min, >8000 RPM: 10 min |
The flash range is 60 to 32,000 RPM/FPM.
The resolution is 0.1 RPM for speeds under 1,000 RPM and 1 RPM for speeds at or above 1,000 RPM.
It accepts AC 110V, 220V, or 230V at 50/60 Hz, or can be powered via an included DC 9V adapter.
Replace the tube only by a qualified technician. Power off the device and wait 15 minutes for discharge. Remove the front cover screws, disconnect the old tube, install the new one, and reassemble the cover.