The manual is the operation manual of SE-309 digital thermometer, which includes the function, use method, technical parameters, safety use and other contents.
This is the user manual for the TestLink SE-309 digital thermometer, a four-channel device designed for use with K-type thermocouples. It provides instructions for power operation, thermocouple connection, temperature scale selection, data-hold, T1-T2 difference mode, memory recording and erasing, clock setup, recording interval, max/min tracking, auto power-off, low battery indication, calibration points, and digital output configuration. The device also includes RS232 communication for PC integration and comes with software for Windows systems.
The thermometer features a 4-digit LCD display per channel, a measurement range of -200°C to 1370°C (-328°F to 2498°F), and internal memory capable of storing 16,000 records per channel. It operates within a temperature range of 0°C to 50°C and includes a 9V battery or AC adapter for power.
| Measurement Range | -200°C ~ 1370°C (-328°F ~ 2498°F) |
|---|---|
| Resolution | 0.1°C from -200°C to 200°C; 1°C from 200°C to 1370°C |
| Accuracy | ±(0.2% reading + 1°C) from -200°C to 200°C; ±(0.5% reading + 1°C) from 200°C to 400°C; ±(0.2% reading + 1°C) from 400°C to 1370°C |
| Display | 4 digital Liquid Crystal Display per channel |
| Memory Capacity | 16,000 records per channel |
| Dimensions | 184× 64 × 30mm |
| Weight | 250g |
| Operating Temperature | 0°C ~50°C (32°F ~ 122°F) |
| Storage Temperature | -10°C to 60°C (14°F ~ 140°F) |
| Sample Rate | 3 seconds per period |
| Digital Output | 9600bps N 81 serial interface |
The measurement range is -200°C to 1370°C (-328°F to 2498°F).
It can store up to 16,000 records per channel.
Accuracy is ±(0.2% reading + 1°C) from -200°C to 200°C, ±(0.5% reading + 1°C) from 200°C to 400°C, and ±(0.2% reading + 1°C) from 400°C to 1370°C.
Calibration points are at 0°C (VR1 ± 0.1°C), 190°C (VR2 ± 0.1°C), 1000°C (VR3 ± 1°C), and 1900°F (VR4 ± 1°F).