Introduces the basic functions and operations of electronic multi-channel pipettes
This manual provides instructions for operating Rainin's EDP3-Plus adjustable-nozzle multichannel pipettes, which are based on single-channel models. It covers nozzle spacing adjustment using LIMITER and SPACING ADJUSTMENT knobs, tip mounting with LTS tips, immersion depth, and auto-dispensing features. The document also includes specifications, care instructions, and contact information for technical support.
The pipettes are available in 6-channel (EA6) and 8-channel (EA8) versions with volume ranges from 5-50 μL to 100-1200 μL. Proper use requires RAININ LTS tips, and the device must be handled with care to maintain precision. Maintenance involves occasional cleaning with distilled water, avoiding wetting the display or battery compartment, and storing the pipette safely when not in use.
| Volume Range | 5-50 μL (EA8-50), 20-300 μL (EA6-300, EA8-300), 100-1200 μL (EA6-1200, EA8-1200) |
|---|---|
| AutoPace Speed Setting | Maximum speed setting: 10 for 50 & 300 μL models, 9 for 1200 μL models |
Use the LIMITER knob to set the maximum spacing, then turn the SPACING ADJUSTMENT knob to open the spacing to the desired value.
EA8-50: 2-3 mm; EA6-300, EA8-300: 3-6 mm; EA6-1200, EA8-1200: 6-10 mm.