This document is the user manual for Wahl DST6xx series thermometers
This manual provides detailed instructions for installing, operating, and calibrating Wahl DST6xx series thermometers. It covers models including DST600, DST610, DST620, DST640, DST650, and DST660, with features such as platinum RTD sensors, self-checking electronics, and probe error detection. The document also includes sections on battery installation, ferrite installation for EMI compliance, and panel mounting options.
The manual is intended for users in food, dairy, and pharmaceutical industries requiring high accuracy and reliability, with optional versions for international shipping, adjustable resolution, and panel mounting.
| Sensor Type | High reliability, 4-wire, 100 ohm, .00385Ω/Ω/°C, thin-film platinum RTD sensor per DIN EN 60751, Class A |
|---|---|
| Accuracy | High accuracy 24-bit Delta-Sigma Analog/Digital Converter |
| Self-Checking Circuitry | Meter repetitively checks the electronics accuracy by measurement of on-board ultra-stable, low ppm precision resistors against embedded firmware values and displays “err” in the event the electronics is out by more than ±0.5°F (0.3°C) |
| Probe Error Checking | Checks for open wire, open sensor, shorted sensor, incorrect wiring |
| Probe ID Feature | Alerts the operator that the probe has been changed and the calibration may be void. Requires the use of calibration software to calibrate and associate the new probe with the meter. Available on fixed stem and remotes less than 25 meters. |
| Programmable R0 | Allows programming of the sensing probes specific R0 value into the meter for accurate temperature calibration |
| Enclosure | DST600 – Temperature Indicating Device – Nema 4X rated Stainless Steel enclosure |
| Optional Versions | “I” version for International shipping using a AA battery, “R” version with adjustable resolution and “P” version for panel mount option |
The DST6xx series incorporates a high reliability, 4-wire, 100 ohm, .00385Ω/Ω/°C, thin-film platinum RTD sensor per DIN EN 60751, Class A, and a high accuracy 24-bit Delta-Sigma Analog/Digital Converter.
The meter repetitively checks the electronics accuracy by measuring on-board ultra-stable, low ppm precision resistors against embedded firmware values and displays “err” if the electronics are out by more than ±0.5°F (0.3°C).
It alerts the operator that the probe has been changed and the calibration may be void. Requires calibration software to associate the new probe with the meter.
The optional versions include “I” for international shipping with AA battery, “R” with adjustable resolution, and “P” for panel mount option.