This manual introduces the EA-31000 Auto CO2 Lab Control Unit, including safety instructions, system setup, operating system, changing system presets, flow rate and preset times.
The Euthanex EA-31000 is a programmable logic controller (PLC) designed to deliver a set volume of CO2 into an individual animal cage through the Euthanex Lid. The unit operates at a fixed flow rate determined by an external CO2 source and automatically shuts off the gas after the euthanasia cycle is completed. It is intended for use by properly certified technicians only.
| Voltage | 100-240 v ac |
|---|---|
| Frequency | 50-60hz |
| Max Current | 1 amp |
| Max Inlet Pressure | 75 psi |
| Operating Temperature | 0 to 40°C |
| Relative Humidity | 10 to 85% RH |
| Default Preset Time | 5 minutes |
| Default Dwell Time | 5 minutes |
Only technicians that are properly certified should operate the system.
No, the system and its components must only be modified or repaired by Euthanex authorized service technicians.
The maximum inlet pressure is 75 psi.
The flow rate is derived by multiplying the chamber volume (Height x Width x Length) by 20%.