This manual introduces the use of the 7500 Real-Time PCR System.
This manual covers the operation of the Blackhills Diagnostic Resources 7500 Real-Time PCR System. It details how to create templates and set up experiments using the desktop software, including defining targets and samples for both quantitation and genotyping assays. The document explains how to configure run methods and visualize results.
An amplification curve is considered positive if the cycle threshold (Ct) is less than 35.
The baseline should be set to start at cycle 3 and end at cycle 18.
The mutated allele is detected in the FAM channel, while the wild type allele is detected in the HEX/VIC channel.