This manual describes the operation and installation methods of DTI MRI
The PPMI DTI Operations Manual Version 7.0 outlines the protocols for acquiring functional magnetic resonance imaging (fMRI) and Diffusion Tensor Imaging (DTI) data. It specifies requirements for baseline scans, including 3D T1-weighted and T2-weighted sequences, and details the data transmission procedures to the imaging core lab.
The baseline scan requires a scanner with at least a 1.5 Tesla field strength.
The total scan time is expected to be in the range from between 20 – 30 min.
The slice thickness must be 1.5 mm or less with no interslice gap.