This is the operation manual of Zutron Automatic Endoscope Leak Tester, which introduces the product's intended use, model, installation method and usage methods.
The Zutron ZUTR10003 Automatic Endoscope Leak Tester is designed for early detection of small holes in the exterior of endoscopes to prevent cross-contamination and costly repairs. It is compatible with endoscopes from major manufacturers including ACMI, Fujifilm, Fuse, Olympus, Pentax, Richard Wolf, and Storz. The device operates using a DC 12V, 3.3A power supply and is intended for indoor use within a temperature range of 10°C to 40°C.
| Voltage | DC 12V, 3.3A |
|---|---|
| Temperature range | 10°C/50°F to 40°C/104°F |
| Calibration pressure | 120-196 mmHg |
It is compatible with ACMI, Fujifilm, Fuse, Olympus, Pentax, Richard Wolf & Storz endoscopes.
The unit must run for at least 30 seconds before placing the endoscope in water.
You must wait 10 seconds after powering the toggle switch off to release the air from inside the endoscope.