This manual describes how to maintain, test, troubleshoot, and repair the LIFEPAK 20 defibrillator/monitor or LIFEPAK 20e defibrillator/monitor (device).
This service manual provides detailed instructions for maintaining, testing, troubleshooting, and repairing the LIFEPAK 20/20e defibrillator/monitor. It includes sections on safety, device description, modes of operation, performance inspection, calibration, troubleshooting, and preventive maintenance. The manual also specifies required test equipment and outlines service personnel qualifications.
The document is designed for qualified service technicians and includes a Performance Inspection Procedure (PIP) checklist with test equipment requirements and step-by-step verification procedures for device functionality.
| Test Equipment | Fluke® Biomedical Corp. 215A/217A Defibrillator analyzer, QED 6™ with test posts accessory, Dale model 600 safety analyzer, Agilent 33120A function generator, Tektronix® 2232 oscilloscope |
|---|---|
| Service Personnel Qualifications | Associate of Applied Science in biomedical electronics, Certificate of Technical Training in biomedical electronics, or equivalent experience |
Required equipment includes Fluke® Biomedical Corp. 215A/217A Defibrillator analyzer, QED 6™ with test posts accessory, Dale model 600 safety analyzer, Agilent 33120A function generator, and Tektronix® 2232 oscilloscope.
Service technicians must have an Associate of Applied Science in biomedical electronics, a Certificate of Technical Training in biomedical electronics, or equivalent experience.