This manual provides instructions for using and maintaining the Morehouse Instrument PM-5250 model hydraulic hand pump and jack.
This manual provides instructions for the Morehouse Instrument PM-5250 Hydraulic Hand Pump & Jack, a precision tool designed for calibration work with a maximum ram travel of one inch and a capacity load pressure of approximately 4500 psi. The system utilizes a dual-volume hand pump with a double-diameter piston and an auxiliary screw piston for applying minute increments of force.
| Maximum Ram Travel | one inch |
|---|---|
| Hydraulic Pressure at Capacity | approximately 4500 psi |
| Pump Design | dual volume design based on a double diameter piston |
| Recommended Fluid Viscosity | ISO Viscosity Grade of 22 or 32 |
The manual recommends petroleum base anti-wear hydraulic fluid with an ISO Viscosity Grade of 22 or 32.
Push the valve rod in to use the reduced diameter piston, which is effective for high pressure with minimum effort.
If the system feels spongy, there may be air in the system, and the trapped air needs to be bled out.