This manual introduces the use of oxygen sensors. It includes unpacking the product, connecting the sensors, calibrating the sensors, measuring the sensors, storing and maintaining the sensors.
This manual provides instructions for using Unisense oxygen microsensors in research settings. It covers unpacking, polarization, calibration, mounting, and maintenance. The sensor is a Clark-type device designed for high spatial resolution measurements in physiology, biotech, and environmental sciences. It requires a picoampere amplifier and must be pre-polarized before use.
Troubleshooting is included for common issues like high drifting signals, low response, or unstable readings, with solutions such as degassing water, replacing the sensor, or grounding the setup.
| Warranty | 180 days limited warranty |
|---|---|
| Polarization Voltage | -0.80 V |
| Pre-Polarization Time | At least 2 hours for new sensors |
| Amplifier Compatibility | Unisense picoampere amplifiers (e.g., UniAmp, O2 UniAmp) |
The oxygen sensor is covered by a 180 days limited warranty.
A new sensor must be pre-polarized for at least 2 hours before calibration and use.
Compatible amplifiers include Unisense picoampere amplifiers such as UniAmp, O2 UniAmp, Microsensor Multimeter, and OXY-Meter.
Replace the oxygen microsensor, as contamination or loss of the gold tip may be the cause.