This document describes the cleaning procedures necessary to change a standard plug-in crystal monitor correctly.
This document outlines the procedure for changing a standard plug-in crystal monitor, specifically for use with evaporators from manufacturers such as Cha, Ultex2, Tescal, and NRC. The process involves removing shields, extracting the old monitor assembly, cleaning the components, and installing a new crystal monitor with a spring clip.
| Crystal Frequency | 6 MHz |
|---|---|
| Crystal Monitor Cost | Under $10 |
| Replacement Threshold | 100% life consumed |
The crystal must be changed when it reads more than 100% life consumed, though changes as early as 80% are appropriate for sensitive processes.
The assembly must be cleaned until the stainless steel surfaces are shiny, using Scotch-Brite, IPA, or a razor blade depending on the level of cleanliness.