This document is the operation manual for CAV 4.2 Dual-Bath Kinematic Viscometer, which introduces the features and characteristics of the product and its usage method.
The document provides comprehensive instructions for operating the CANNON CAV 4.2 Dual-Bath Kinematic Viscometer. It details procedures for unpacking, physical setup, and configuring the device via its graphical user interface (GUI) or VISCPRO II software. Key topics include setting bath temperatures, calibrating sensors, managing wash cycles, and running sample tests to measure kinematic viscosity according to ASTM D445/D446 and ISO 3104/3105 standards.
| Dimensions (W x D x H) | 36 cm x 66 cm x 72 cm (14.25 in x 26 in x 28.5 in) |
|---|---|
| Weight | 63 kg (140 lb) |
| Maximum throughput | 24 tests per hour |
| Automated sample capacity | 28 positions (2 x 14 positions) |
| Minimum sample/solvent volume required | 8 mL sample/15 mL solvent per test |
| Viscosity range | 0.5 mm²/s to 10,000 mm²/s in 100-fold increments |
| Timing resolution | 0.01 seconds (timing accuracy to + 0.001 seconds) |
| Power requirements | 100 Vac, 50/60 Hz, 1200 W; 115 Vac, 50/60 Hz, 1200 W; 230 Vac, 50/60 Hz, 1200 W |
The instrument complies with ASTM D445/D446, ISO 3104/3105, and IP 71 methodologies.
The unit operates from 20 °C to +100 °C (+ 0.01 °C), with extended ranges available using integrated coolers or high-temperature baths.
The CAV 4.2 features an automated sample capacity of 28 positions, arranged as 2 x 14 positions.
The device can be controlled using its built-in GUI or via the VISCPRO II software interface.