This manual describes the features and characteristics of the DM-100 series sensors
The Teledyne Detcon Model DM-100 is a non-intrusive, 4-20mA loop-powered sensor designed to detect a wide range of toxic gases and monitor oxygen deficiency. The basic assembly consists of an intrinsically safe electrochemical fuel cell mounted in an explosion-proof stainless steel housing. The sensor is field-proven and features a long service life of up to three years for toxic gases and 2.5 years for oxygen deficiency sensors.
The manual provides detailed instructions for installation in hazardous locations, including sensor placement and mounting guidelines. It covers electrical installation, field wiring, and initial startup procedures for both toxic gas and O2 deficiency sensors. The document also details operation modes, including auto span level adjustment and calibration procedures (zero and span). Additionally, it includes a troubleshooting guide for smart display error codes and service/maintenance instructions for replacing plug-in sensors and terminal boards.
| Power Supply | 4-20mA loop powered |
|---|---|
| Toxic Gas Range | 0-1ppm to 0-10,000ppm |
| O2 Deficiency Range | 0-1% to 0-25% by volume |
| Operating Temperature | -20° to 50°C (-4° to +122° F) |
| Warranty Period | Two years |
| Service Life (Toxic) | Up to three years |
| Service Life (O2) | Up to 2.5 years |
| Electrical Classifications | Class I, Division 1, Groups B, C, and D |
The toxic gas sensors have a long service life of up to three or more years.
The O2 deficiency sensors have an expected service life of up to 2.5 years in ambient air at 20.9% oxygen.
The sensor is warranted for two years.
Yes, the sensor is designed for hazardous locations with electrical classifications of Class I, Division 1, Groups B, C, and D.