This manual introduces general information, installation, operation, principles of operation, maintenance, schematics, parts list, radiation survey, maintenance schedule, etc. for Faxitron Cabinet X-ray Systems Models 43855A, 43855B and 43856A.
Faxitron models 43855A, 43855B, and 43856A are self-contained, radiation-shielded cabinet X-ray systems designed for high resolution radiographs of small to medium-size objects. The cabinets and tables limit X-ray leakage to 0.5 mR/hour or less at 5 cm from any external surface, and the door has dual interlocks that prevent X-ray generation when open. The standard self-rectifying thermionic X-ray tube has a beryllium window and a 0.5 mm focus rated at 110 kV max; optional tube configurations are also listed. Manual controls select exposure time and tube kilovoltage, while option A02 and A02M adds automatic exposure control using an ionization chamber under the film shelf.
This technical manual, manual no. 5081-9519, dated Jan. 1, 1996, covers general information, installation, operation, principles of operation, maintenance, schematics, part list, radiation survey, and maintenance schedule. It includes safety warnings that maintenance is to be performed by qualified personnel only and that radiation surveys are required before operation and after moving or servicing the system.
| X-ray leakage limit | 0.5 mR/hour or less at 5 cm from any external surface |
|---|---|
| Standard X-ray tube focus | 0.5 mm |
| Standard X-ray tube maximum voltage | 110 kV max |
| Optional X-ray tube configurations | 0.5 mm focus/130 kV; 0.07 mm focus/110 kV; 0.05 mm focus/55 kV |
| Cabinet door interlocks | Dual interlocks prevent X-ray generation when the door is open |
Only qualified personnel may service the system; serious physical and genetic injury can result if all X-ray shielding is not properly replaced or interlocks are not operating after maintenance.
A radiation survey is required before placing the system in operation, before use whenever the system is moved or serviced, during installation, after extensive maintenance, and periodically not to exceed six months.
A Victoreen Model 450 P or equivalent is required; Geiger-Mueller and certain other scintillator-type radiation meters are not acceptable.
The manual covers Faxitron Cabinet X-ray Systems models 43855A, 43855B, and 43856A.