Faxitron 43855A 43855B 43856A Cabinet X-ray Systems Technical Manual

This manual introduces general information, installation, operation, principles of operation, maintenance, schematics, parts list, radiation survey, maintenance schedule, etc. for Faxitron Cabinet X-ray Systems Models 43855A, 43855B and 43856A.


Faxitron 43855A 43855B 43856A Cabinet X-ray Systems Technical Manual - cover page
Brand
Faxitron
Category
Other
Document type
Service Manual
Language
English
Pages
111
File format
PDF
File size
1.1 MB
Published
16 October, 2023
Updated
16 October, 2023
MD5 checksum
4A80E281E3CF5F732BFC3F8E85A8ADCA

About this manual

Faxitron models 43855A, 43855B, and 43856A are self-contained, radiation-shielded cabinet X-ray systems designed for high resolution radiographs of small to medium-size objects. The cabinets and tables limit X-ray leakage to 0.5 mR/hour or less at 5 cm from any external surface, and the door has dual interlocks that prevent X-ray generation when open. The standard self-rectifying thermionic X-ray tube has a beryllium window and a 0.5 mm focus rated at 110 kV max; optional tube configurations are also listed. Manual controls select exposure time and tube kilovoltage, while option A02 and A02M adds automatic exposure control using an ionization chamber under the film shelf.

This technical manual, manual no. 5081-9519, dated Jan. 1, 1996, covers general information, installation, operation, principles of operation, maintenance, schematics, part list, radiation survey, and maintenance schedule. It includes safety warnings that maintenance is to be performed by qualified personnel only and that radiation surveys are required before operation and after moving or servicing the system.

What's inside

  • General Information
  • Installation
  • Operation
  • Principles of Operation
  • Maintenance
  • Schematics
  • Part List
  • Radiation Survey
  • Maintenance Schedule

Specifications

X-ray leakage limit0.5 mR/hour or less at 5 cm from any external surface
Standard X-ray tube focus0.5 mm
Standard X-ray tube maximum voltage110 kV max
Optional X-ray tube configurations0.5 mm focus/130 kV; 0.07 mm focus/110 kV; 0.05 mm focus/55 kV
Cabinet door interlocksDual interlocks prevent X-ray generation when the door is open

Frequently asked questions

Who is allowed to service the Faxitron system?

Only qualified personnel may service the system; serious physical and genetic injury can result if all X-ray shielding is not properly replaced or interlocks are not operating after maintenance.

When must a radiation survey be performed?

A radiation survey is required before placing the system in operation, before use whenever the system is moved or serviced, during installation, after extensive maintenance, and periodically not to exceed six months.

What radiation meter is required for checking leakage?

A Victoreen Model 450 P or equivalent is required; Geiger-Mueller and certain other scintillator-type radiation meters are not acceptable.

What models are covered by this manual?

The manual covers Faxitron Cabinet X-ray Systems models 43855A, 43855B, and 43856A.