This document is the manual of KRT-01-3PN digital ground resistance tester
The Total Ground KRT-01-3PN is a digital ground resistance meter designed to measure earth electrode resistance, soil resistivity, AC/DC voltage, and low resistance. It features a large LCD display with double imaging, automatic zero adjustment, and an indicator for exceeding limits.
This manual provides instructions on safety precautions, battery replacement, and various measurement methods including the standard three-pole method and simplified two-wire method for high resistances. This manual is written in Spanish.
| Ground Resistance Range | 200, 2000, 20000 Ohms |
|---|---|
| Sampling Rate | 2.5 times per second |
| Operating Temperature | 0°C to 40°C (32°F to 104°F) |
| Power Source | 6x1.5V AA batteries or equivalent |
It is recommended when measuring a ground resistance higher than 10 Ohms and auxiliary earth spikes cannot be driven.
Replace batteries when the low battery symbol appears on the LCD screen.