This is a maintenance manual which includes maintenance instructions, troubleshooting and error messages for Mercury Analyzer MA-3 Solo
This service manual is intended for engineers with proper training who will conduct inspections, repairs, or maintenance of the Mercury Analyzer MA-3 Solo. It provides safety instructions regarding installation environments, power specifications, and handling hot components like the sample inlet and mercury collector tube.
The document details preventive maintenance procedures, including component replacement and washing absorption cells. It also outlines troubleshooting steps for issues such as sensitivity fall and abnormal blank values, along with descriptions of gas flow paths during atomization and measurement steps.
| Measurement Principle | thermal decomposition mercury analyzer |
|---|---|
| Heater H4 Temperature | 150°C |
This service manual is intended for engineers who will have proper training and conduct the inspection, repairs, or maintenance of the Analyzer.
Wear protective equipment such as heat-resistant gloves because the area around the sample inlet becomes hot.