This manual introduces the assembly, operation, maintenance and maintenance of Sunnex Tristar mobile and wall lights.
The document serves as a service and installation guide for the Sunnex Tristar series of medical examination and surgical lights. It details the contents of shipping cartons, assembly procedures for both mobile (Hand-Ez upright column) and wall-mounted versions, and specific installation requirements such as stud support and lag bolt specifications.
| Power Supply | 120V or 230V, 50-60Hz |
|---|---|
| Nominal Effect | 105VA |
| Light Intensity (35W Model) | 38,000 Lux @ 1 meter (with UV filter) |
| Light Intensity (20W Model) | 27,000 Lux @ 1 meter (with UV filter) |
| Light Source (35W Model) | Halogen Bulb –12V/3x35W/10° |
| Light Source (20W Model) | Halogen Bulb –12V/3x20W/10° |
| Color Temperature | 3,000K |
| Weight (Mobile) | 25 lbs |
| Weight (Wall) | 20.4 lbs |
| Medical Standard | Class II |
No, using a wall dimmer voids the certification listing of the product.
Bolts must support a minimum of 100 pounds pull and 100 shear on each bolt.
Use no smaller than 1/2" lag bolts that penetrate wooden wall studs to at least a depth of 2 1/2".
The lamp head is designed to accept a sterilizable handle, which can be ordered separately.