This is a user guide that introduces the installation and operation of the 8200 bubble gauge
This document serves as a user's guide for the Sutron 8200 digital bubbler water level gauge, designed to replace traditional analog chart-recording gauges. It details the system components, including the data collection platform, control unit, battery, and various interface connections for nitrogen supply and orifice tubing.
The manual provides comprehensive instructions on system transportation, installation, tie-down procedures, and making external connections using specific wrench sizes. It also covers start-up procedures, site visits, direct connection data retrieval methods, and guidelines for operation with GOES capability.
| Maximum water height above orifice | 34 feet |
|---|---|
| Pressure sensor limit | 15 psig |
Connections require 7/16, 9/16, 11/16, and 1 inch open wrenches, an adjustable wrench, and a screwdriver.
For new installations, tighten the fittings 1 1/4 turns past finger tight.
Heights greater than 65 feet will cause permanent sensor damage.