Explains how to use Helix Technology Corporation's On-Board Cryopump Helium Circuit Decontamination Kit P/N 8080250K003 to remove gaseous contaminants from the On-Board Cryopump helium circuit.
This document provides instructions for the Helix Technology On-Board Cryopump Helium Circuit Decontamination Kit (P/N 8080250K003). The procedure removes gaseous contaminants from the helium circuit by freezing them in the coldhead of the On-Board Cryopump. This process addresses operational issues such as ratcheting noise and performance degradation caused by contamination.
| Required Helium Purity | 99.999% |
|---|---|
| Regulator Range | 0-400/0-3000 psig |
| Estimated Time for Sequential Decontamination | 45 minutes for first, 30 minutes for each additional |
Ratcheting noise is typically caused by gaseous contamination in the helium circuit, which degrades performance.
Separate decontamination is only required if the compressor has been opened to atmosphere or the helium pressure has dropped to zero.
It takes 45 minutes for the first unit and 30 minutes for each additional unit.