Helix Technology On-Board Cryopump Helium Circuit Decontamination Instructions

Explains how to use Helix Technology Corporation's On-Board Cryopump Helium Circuit Decontamination Kit P/N 8080250K003 to remove gaseous contaminants from the On-Board Cryopump helium circuit.


Helix Technology On-Board Cryopump Helium Circuit Decontamination Instructions - cover page
Brand
Helix Technology
Category
Other Electronics
Document type
Service Manual
Language
English
Pages
14
File format
PDF
File size
113 KB
Published
16 October, 2023
Updated
16 October, 2023
MD5 checksum
8BC64A946757544404C6819DCD42AF12

About this manual

This document provides instructions for the Helix Technology On-Board Cryopump Helium Circuit Decontamination Kit (P/N 8080250K003). The procedure removes gaseous contaminants from the helium circuit by freezing them in the coldhead of the On-Board Cryopump. This process addresses operational issues such as ratcheting noise and performance degradation caused by contamination.

What's inside

  • Purpose
  • Kit Contents
  • Helium Circuit Contamination
  • Methods of Decontamination
  • Equipment/Tools Requirements
  • Decontamination Flowchart
  • Method 1 - Sequential Decontamination

Specifications

Required Helium Purity99.999%
Regulator Range0-400/0-3000 psig
Estimated Time for Sequential Decontamination45 minutes for first, 30 minutes for each additional

Frequently asked questions

What causes ratcheting noise in an On-Board Cryopump?

Ratcheting noise is typically caused by gaseous contamination in the helium circuit, which degrades performance.

When is separate decontamination of the compressor required?

Separate decontamination is only required if the compressor has been opened to atmosphere or the helium pressure has dropped to zero.

What is the estimated time for sequential decontamination of all On-Board Cryopumps?

It takes 45 minutes for the first unit and 30 minutes for each additional unit.