This manual contains information on cleaning, maintenance, troubleshooting and parts for the 503 3-section Couch models.
The document provides comprehensive instructions for servicing the Model 503 3-section couch, including cleaning procedures, decontamination protocols using hypochlorite solutions, and a detailed service checklist. It outlines general warnings regarding load limits (250kg) and electrical safety before performing maintenance tasks.
| Maximum Working Load (SWL) | 250kg distributed evenly |
|---|---|
| Overall Length | 184 cm |
| Height Range | 47-98 cm |
| System Voltage | 24v |
| IP Rating | IPX4 |
It is recommended that all products are serviced annually, while electrically operated products or those in high use should be serviced every 6 months.
To decontaminate the couch, repeat cleaning steps using a dilute solution of Hypochlorite and warm water (10,000ppm).
The correct plug fuse fitted must be 5A.