This manual describes the use of the Newton Scientific EVO 10 Floor Centrifuge
The Newton Scientific EVO 10 Floor Centrifuge is a laboratory product designed to separate substance mixtures of different densities. It must be operated by trained specialists only and requires setup on a horizontally levelled, rigid surface with adequate load-bearing capacity in a well-ventilated environment.
This instruction manual covers the entire lifecycle of the device, including preparation, operation (switching on/off, rotor installation, program selection), maintenance and care, troubleshooting for common errors like imbalance or overspeed, and safety procedures such as mechanical emergency door release during power failures.
| Max RPM | 10,000 rpm |
|---|---|
| RPM accuracy | ±20 rpm |
| Max RCF | 18,300 × g |
| Max Capacity | 6×1000 ml |
| Temp Range | -20℃~40℃ |
| Power Supply | AC 200-240V, 50Hz, 19.6A |
| Sound Level | ≤ 65dBa |
The error code E01 indicates an imbalance, while E02 indicates an overspeed condition.
Ensure the rotor has stopped and the isolator is off, then pull the release cord located on the bottom side of the panel.
Check the loading of the rotor and ensure it is loaded symmetrically to correct any imbalance.