This file is about the maintenance manual of AccuStaltic peristaltic pump, including the regular inspection and maintenance.
The document outlines procedures for periodic inspection and maintenance of AccuStaltic peristaltic pumps. It recommends weekly visual inspections of the rotor and tubing, cleaning with wipes, and re-lubricating with silicone-based or compatible lubricants to maximize service life and accuracy. It warns against pressure-washing to prevent seal damage.
A weekly inspection schedule is recommended for frequent or continuous use.
Silicone-based spray lubricants or greases are recommended for most commercially available tube types.
No, the rotor is not field-serviceable and must be replaced as a complete unit.