Buck Research Instruments CR-3 Hygrometer Operating Manual

This manual introduces the working principle, usage, maintenance and troubleshooting of the CR-3 temperature/humidity sensor.


Buck Research Instruments CR-3 Hygrometer Operating Manual - cover page
Brand
BUCK RESEARCH INSTRUMENTS
Document type
User Manual
Language
English
Pages
26
File format
PDF
File size
404 KB
Published
15 October, 2023
Updated
15 October, 2023
MD5 checksum
AC7E22B8237583EC02C07F17D438467F

About this manual

The Model CR-3 is a high-performance chilled mirror hygrometer designed for laboratory use or process control. It measures dew/frost point temperature by cryogenically cooling a mirror to condense water vapor, utilizing optical detectors and an ultra-stable thermistor for accurate readings without periodic recalibration.

This manual provides instructions on the instrument's components, installation procedures, power-up and power-down sequences, and operation modes including signal processing. It also details maintenance tasks such as sample line cleaning, mirror cleaning, leak testing, and cryocooler checks, along with troubleshooting guides.

What's inside

  • Introduction
  • Specifications
  • Location of Principle Components
  • Installation and Operation
  • Signal Processing
  • Principles of Operation
  • Measurement Limitations
  • Maintenance and Troubleshooting

Specifications

Dew/frost point temperature range-120°C to +30°C
Moisture concentration range<1 ppbv to 3%
Frost point reading accuracy (above -70°C)+0.15°C
Response timeLess than 40 sec typical
Ambient operating temperature range-20 to +40°C
Operating pressure range0 — 150 psia (0 — 10 bar)
Sample flow rate range0.2 - 3 liters/minute
Cryocooler typeKleemenko cycle, mixed refrigerant
Operating lifetimeApprox. 100,000 hours
Input voltage (typical)110VAC 60 Hz

Frequently asked questions

Can the CR-3 be used in aircraft applications?

No, the instrument should never be installed in an aircraft application as this will void the warranty.

What is the typical response time of the CR-3?

The typical response time for the instrument is less than 40 seconds.

How often should the air intake be cleaned?

The air intake should be cleaned every 3 months to prevent overheating.