This manual introduces the working principle, usage, maintenance and troubleshooting of the CR-3 temperature/humidity sensor.
The Model CR-3 is a high-performance chilled mirror hygrometer designed for laboratory use or process control. It measures dew/frost point temperature by cryogenically cooling a mirror to condense water vapor, utilizing optical detectors and an ultra-stable thermistor for accurate readings without periodic recalibration.
This manual provides instructions on the instrument's components, installation procedures, power-up and power-down sequences, and operation modes including signal processing. It also details maintenance tasks such as sample line cleaning, mirror cleaning, leak testing, and cryocooler checks, along with troubleshooting guides.
| Dew/frost point temperature range | -120°C to +30°C |
|---|---|
| Moisture concentration range | <1 ppbv to 3% |
| Frost point reading accuracy (above -70°C) | +0.15°C |
| Response time | Less than 40 sec typical |
| Ambient operating temperature range | -20 to +40°C |
| Operating pressure range | 0 — 150 psia (0 — 10 bar) |
| Sample flow rate range | 0.2 - 3 liters/minute |
| Cryocooler type | Kleemenko cycle, mixed refrigerant |
| Operating lifetime | Approx. 100,000 hours |
| Input voltage (typical) | 110VAC 60 Hz |
No, the instrument should never be installed in an aircraft application as this will void the warranty.
The typical response time for the instrument is less than 40 seconds.
The air intake should be cleaned every 3 months to prevent overheating.