This document is about the operation and use instructions of K16500 series rapid flash point tester
This document is the Technical Manual for the &koeiler K16500 Series Rapid Flash Point Tester. It provides instructions on operating the instrument to determine the flash point of volatile materials using equilibrium closed cup technology compliant with ASTM, IP, ISO, and BSI standards.
The manual covers preparation for use including assembly, detailed operation procedures for both unknown flash point determination and Flash/No Flash verification tests, and a maintenance guide. It also includes an appendix with parts lists, sample scales, assembly drawings, and service information.
| Flash Point Temperature Range | -36°F to 572°F (-30°C to 300°C) |
|---|---|
| Electrical Power | 98-132v, 50/60 Hz or 196-264v, 50/60 Hz |
| Cup Type | Closed, equilibrium test |
| Operational Sample Size (<=212°F) | 2 ml |
| Operational Sample Size (>212°F) | 4 ml |
| Physical Dimensions | 15.0 x 3.4 x 5.3 inches (38.1 x 8.64 x 16.25 cm) |
| Net Weight | 10 lbs (4.6 Kg) |
The test cup and lid dimensions meet ASTM D 3243, D 3278, D 3828; IP 303; ISO 3679, 3680; and BSI BS 3900 Parts A11, A14 requirements.
A sample size of 4 ml is required for testing at temperatures above 212°F (100°C).
No, the Refrigerant Charged Cooling Block K16509 is an optional accessory required for below ambient testing.