This document is a product manual for a gas detector
The SENSIT GOLD is a gas detection instrument designed to detect combustible gases, oxygen content, and toxic gases when equipped with available sensors. It features an advanced low power semiconductor sensor, an automatically backlit display, and audible and visual alarms for hazardous conditions. The instrument is approved for use in Class I, Division 1, Groups C & D hazardous locations when used with approved batteries.
This manual provides instructions for preparation, safety precautions, parts identification, general description, and detailed operation. It covers battery installation, user menu functions, calibration procedures (including Smart-Cal), printing logs, and warranty information. The document includes specifications for various sensors (LEL, CO, O2, H2S) and product dimensions.
| LEL Range | 0-100% LEL |
|---|---|
| CO Range | 0-2000ppm |
| H2S Range | 0-100ppm |
| Size | 11.5" x 3" x 2.32" (292 x 76 x 59 mm) |
| Operational Temp | 0 to 120° F (-17.8 to 48.8° C) |
| Storage Temp | -20° to 132° F (-28.9 to 55.6° C) |
| Battery Life (Alkaline) | 12 hrs. continuous |
The SENSIT GOLD is designed to detect combustible gases, oxygen content, and toxic gases when so equipped with the available sensors.
The combustible gas alarm is preset at 10% LEL methane.
The carbon monoxide (CO) alarm is factory preset at 50ppm.
The instrument is warranted to be free from defects in materials and workmanship for a period of two years after purchase.