This manual introduces the maintenance and repair of electronic tube test sets TV-2/U, TV-2A/U, TV-2B/U and TV-2C/U, including operation instructions, maintenance methods, fault troubleshooting methods, etc.
This document serves as the Field and Depot Maintenance Manual for the Test Sets, Electron Tube TV-2/U, TV-2A/U, TV-2B/U, and TV-2C/U. It covers maintenance functions allocated to the fourth and fifth echelons, including troubleshooting, testing, aligning, and repairing the equipment. The manual also lists the necessary tools, materials, and test equipment required for these maintenance tasks.
| Filament Voltage Range | 0.625 to 117 volts alternating current |
|---|---|
| Maintenance Echelon | Fourth and Fifth Echelon |
Users must be careful when working on the 290-volt plate and screen supply circuits or the 115-volt ac line connections, as serious injury or death may result.
This manual provides instructions for field and depot maintenance, including troubleshooting, testing, aligning, repairing, and replacing parts for Test Sets, Electron Tube TV-2/U, TV-2A/U, TV-2B/U, and TV-2C/U.
Be careful when working on the 290-volt plate and screen supply circuits.
The filament supply provides 19 different filament voltages ranging from 0.625 to 117 volts alternating current.