This manual provides information on the operation, use, precautions and maintenance of the Valco Instruments Co. Inc. Thermal Conductivity Detector (TCD-3).
The Valco Instruments TCD-3 is a thermal conductivity detector designed to operate as a standalone unit or as part of a larger chromatographic system. It features a unique dual-filament design with two separate filaments (A and B) and a controller module that incorporates electrometer and temperature controls. The system provides analog output signals for each channel and a referenced differential output, maintaining thermal stability within 0.010°C to produce a stable, low-noise signal.
| Thermal Stability | within 0.010°C |
|---|---|
| Recommended Carrier Flow Rate | 1 – 10 mL/min |
| Analog Output Signal Range (10V) | -10V to +10V |
| Analog Output Signal Range (1V) | -1V to +1V |
| Installation Category | Installation (overvoltage) II, pollution degree 2 |
| Power Supply Rating | 2A, 250V |
Yes, the TCD-3 is designed to operate either as a standalone test unit or as a component in a larger chromatographic system.
It operates by measuring the change in current required to maintain a constant filament temperature as gases of varying thermal conductivities cross the filament.
The exterior should be cleaned with a dusting brush, or with a moistened cloth (99% water + 1% mild detergent) if necessary.
No, it is approved for use only in heavy industrial environments and may not be used in residential, commercial, or light-industrial environments.