This document is ESA601 Electrical Safety Analyzer Getting Started Manual, published by Fluke Biomedical company, which includes product overview, product features, product parameters, installation instructions, operation instructions, maintenance instructions, troubleshooting and so on
This document serves as the Getting Started Manual for the Fluke Biomedical ESA601 Electrical Safety Analyzer. It provides instructions on unpacking the unit, selecting language and printer options, connecting the Device Under Test (DUT), and performing various electrical safety tests. The manual also details error codes and warranty information.
| Operating Voltage Range | Minimum 90 V AC Maximum 264 V AC |
|---|---|
| Size | 9.8 L x 3 D x 8.25 W (inches) |
| Weight | ≤5 pounds |
| Operating Temperature | 10 to 40 °C |
| Storage Temperature | -25 to 50 °C |
| Maximum Humidity | 80 % relative humidity up to 31 °C |
| Leakage-Current Measurement Range | 0 8000 µA |
| Voltage Measurement Range | 90 to 264 V |
| Earth-Resistance Measurement Range | 0 to 1.999 Ω |
| Insulation Measurement Range | 0.5 MΩ to 100 MΩ |
| VDE Differential Current Range | 10 µA to 10000 µA |
| Mains on Applied Parts Voltage | ≥ 110 % of input Mains voltage |
Fluke Biomedical warrants the instrument against defects in materials and workmanship for one full year from the date of original purchase.
The ESA601 should be calibrated annually.
No, there are no user replaceable parts and the product should not be opened.
The baud rate is 9600 with 1 start bit, 8 parity bits, and 1 stop bit.