This document is the K series chiller instruction manual, which introduces the characteristics of the chiller, installation, operation, maintenance and warranty.
This document serves as the official instruction manual for the Applied Thermal Control K1 and K3 series standard models of chillers. It provides comprehensive guidance on unpacking, site requirements, installation procedures including voltage selection and fluid filling, and operational controls using the built-in PID controller.
The guide also details maintenance schedules, service requirements, status light interpretation, and troubleshooting steps for common issues. Additionally, it outlines warranty registration processes, return of goods procedures, and technical specifications regarding dimensions and performance.
| Controller accuracy | within 0.1°C of set point |
|---|---|
| Voltage (K1) | 230V +10% (50Hz) single phase |
| Voltage (K3) | 230V +10% (50Hz) single phase |
| Circuit breaker rating (K1) | 10 Amps |
| Normal operating current (K1) | 6Amps |
| Circuit breaker rating (K3) | 15 Amps |
| Normal operating current (K3) | 13Amps |
| Ambient temperature range | +4°C to +40°C |
ATC provides a comprehensive return to base 2-year parts and 1-year labour warranty from delivery as standard.
Yes, switch selectable options for operation at 208VAC 60Hz, 230VAC 50Hz and 220VAC 60Hz are available on request.
Hexid fluids are the preferred coolant choice as they provide excellent corrosion protection, freeze protection, algae inhibition and good heat transfer properties.
Push the 'P' button to flash 'SP 1', then press 'UP' or 'DOWN' buttons to adjust. Push 'P' again to store after approximately 10 seconds.