Sutter Instrument Lambda DG-4 and DG-5 Operation Manual

This manual is the operation manual of the Lambda DG-4 and Lambda DG-5 ultra-high-speed wavelength switching illumination system. It covers the use, maintenance and troubleshooting of the system.


Sutter Instrument Lambda DG-4 and DG-5 Operation Manual - cover page
Brand
Sutter Instrument
Category
Other Electronics
Document type
User Manual
Language
English
Pages
82
File format
PDF
File size
2.9 MB
Published
12 October, 2023
Updated
12 October, 2023
MD5 checksum
5D4715BFB97248A41D245DA474EAF045

About this manual

This manual provides operating instructions for the Lambda DG-4 and Lambda DG-5 ultra-high-speed wavelength switching illumination systems. These instruments are designed specifically as microscope illuminators and optical wavelength switchers for use on animal tissues in a laboratory environment. The document details the system's components, including the xenon arc lamp and optional SmartShutter, and outlines proper usage to ensure safety and performance.

What's inside

  • Safety Warnings
  • Electrical Specifications
  • Fuse Replacement
  • Operational Precautions
  • Xenon Bulb Replacement
  • Filter Installation
  • Technical Specifications
  • Warranty Information

Specifications

Input Voltage110-120 V AC, 60 Hz, or 220-240 V AC, 50 Hz
Power Fuse Rating (175W)4.0A, 125V
Power Fuse Rating (300W)5.0A, 125V
Product Versions175W and 300W versions

Frequently asked questions

Is the Lambda DG-4/5 intended for human experimentation?

No, the instrument is designed for use on animal tissues and is not a medical device.

What type of fuse should be used for the 175W version?

A 4.0A, 125V time delay (medium) or slow blow glass tube fuse.

Does the full spectrum bulb generate ozone?

Yes, it generates significant amounts of ozone, requiring a ventilation or ozone-removing system.

Download PDF (2.9 MB)

More Sutter Instrument Manuals


View all Sutter Instrument manuals →