This document introduces the instruction manual of ADH® NETCOM™ automatic air dehydrator, including safety information and warnings, installation instructions, operation instructions and maintenance instructions.
This manual provides comprehensive instructions for the Environmental Technology ADH NETCOM Automatic Air Dehydrator. It covers safety information, unpacking, inventory checks, and detailed setup procedures including machine preparation and installation. The document explains the system's theory of operation, computer network communications, and how to configure user inputs and alarms.
| Maximum operating temperature | 130°F (55°C) |
|---|---|
| Air pump pressure capability | 24 psig (1,655mbar) |
| High Limit Target Pressure range | 0.20 psig to 7.5 psig (14mbar to 517mbar) |
| High Pressure Alarm Level default | 1.5 psig (103.5mbar) |
| Low Limit Target Pressure default | 0.30 psig (21mbar) |
| Low Pressure Alarm Level default | 0.15 psig (10mbar) |
| High Duty Cycle Alarm Level default | 50% |
Immediately interrupt power to the ADH NETCOM with the POWER switch at the rear of the unit, unplug the unit, or turn off the circuit breaker controlling the outlet.
The maximum operating temperature is 130°F (55°C).
The factory default setting for High Limit Target Pressure is 0.5 psig (34.5mbar).