This product manual introduces the features and characteristics of L3, L3S, 757, 765, L6, and L6S spectral series products.
This document is the user manual for the INESA Spectral Series, covering the L3, L3S, 757, 765, L6, and L6S visible and ultraviolet-visible spectrophotometers. It explains the operating principles based on colorimetry and the Beer-Lambert law, detailing the instrument's features such as the large color touch screen and USB data interface. The guide is intended for qualitative and quantitative analysis in laboratory environments. This manual is written in Chinese.
| Wavelength Range | 325nm~1100nm (L3, L3S, 757); 190nm~1100nm (765, L6, L6S) |
|---|---|
| Optical System | Single beam with monitor light path, 1200 lines/mm holographic grating system |
| Wavelength Accuracy | ±0.5nm |
| Wavelength Repeatability | ≤0.2nm |
| Transmittance Range | 0.0%~200.0% |
| Absorbance Range | -0.301A~4.000A |
| Minimum Spectral Bandwidth | 2±0.4nm |
| Receiver Element | Silicon photovoltaic cell |
| Screen Size | 8 inch |
| Weight | 20kg |
| Power Supply | AC220V±22V, 50Hz±1Hz |
The maximum allowable error for the wavelength is ±0.5nm.
The wavelength should be calibrated once or twice a year, or after replacing the light source.