This document is a user manual for the battery-powered portable digital insulation resistance meter KEW3551/3552/3552BT. The document includes safety warnings, features, specifications, component names, accessories, preparation for testing, voltage testing, insulation resistance testing, good/bad judgment function, DAR/PI testing, low resistance testing, backlight/LED light and buzzer function, and automatic power off function.
The KEW 3551/3552/3552BT is a battery-powered digital instrument designed for voltage testing, insulation resistance testing, and low resistance testing. It features an automatic range selection, a built-in LED light for low-light conditions, and a buzzer function. The KEW3552 and KEW3552BT models offer additional capabilities such as DAR/PI testing, memory storage, data transmission via cable, and Bluetooth communication for remote control and data saving on tablets. This manual is written in Chinese.
| Test Categories | CAT IV 300V / CAT III 600V |
|---|---|
| Voltage Test Range | AC 300.0V / 600V, DC 300.0V / 600V |
| Insulation Resistance Range | 4MΩ to 40GΩ (depending on model) |
| Low Resistance Test Range | 0.20Ω to 4000Ω |
| Power Source | 4 AA Alkaline Batteries |
| Dimensions | 97 x 156 x 46 mm |
| Weight | Approx. 490g (with batteries) |
| Bluetooth Version | Ver 4.0 |
The instrument is designed for CAT IV 300V and CAT III 600V environments.
The instrument has an automatic discharge function that discharges charges in capacitive loads after testing.
Only the KEW3552BT model supports Bluetooth wireless communication.
The power is automatically cut off after 10 minutes of no operation.