This document introduces the main functions, technical parameters, installation, and operation methods of the XK3150-Ex intrinsically safe electronic weighing instrument. The instrument has powerful functions, excellent indicators, and reliable quality, suitable for weighing requirements in explosive gas environments and combustible dust environments.
The XK3150-Ex is a intrinsically safe electronic weighing instrument designed for explosive gas and combustible dust environments. It complies with GB3836 standards and offers two power options: a certified explosion‑proof battery pack or an intrinsically safe output power supply.
The manual covers safety precautions, installation and wiring, basic functions such as weighing, zeroing, tare and counting, calibration procedures, default parameter settings, error handling, routine maintenance and the complete packing list. This manual is written in Chinese.
| Resolution | 30,000 |
|---|---|
| Zero drift | < 0.1 μV/K |
| Measurement speed | 50 ms |
| Signal range | -30~30 mV DC |
| Bridge supply voltage | 5 VDC |
| Power supply options | AC 220V, DC12V (output) or 7.4V/4Ah battery |
Only qualified personnel may set up, debug, test or repair the instrument; ensure reliable grounding; disconnect power before any electrical connection; avoid touching internal components and use anti‑static measures.
When the display shows “LOBATT”, remove the four screws, take out the battery to a safe area, charge it; the charger’s red light indicates charging and the green light indicates full charge.