This document provides detailed information about the principles, uses, and characteristics of the 721/721N visible spectrophotometer. It also includes specifications and standard accessories of the instrument. The document provides a guide for operating the instrument, including installation conditions and unboxing inspection. Additionally, the document covers basic operation and preheating process of the instrument.
The document serves as a user guide for the Hengping 721/721N Visible Spectrophotometer. It details the instrument's specifications, including its wavelength range of 335 to 1000nm and optical system components like the holographic grating and halogen lamp source.
The manual covers various operational modes such as transmittance, absorbance, concentration factor, and direct reading concentration. It provides step-by-step instructions for setting up the instrument, performing zero and 100% adjustments, calibrating wavelength accuracy using magnesium oxide glass, and troubleshooting common faults. This manual is written in Chinese.
| Wavelength Range | 335 ~ 1000nm |
|---|---|
| Optical System | C-T type monochromator, 1200 lines/mm holographic grating |
| Light Source | Halogen lamp (COW/1 12V) |
| Wavelength Accuracy | +/- 2nm |
| Photometric Accuracy | +/- 1.0%(T) |
| Power Supply | 220V+10% 50Hz |
The instrument operates within a wavelength range of 335 to 1000nm.
The manual recommends a preheat time of not less than 30 minutes to reach thermal equilibrium.
To calibrate, set the display to 800nm with a blank sample, then measure magnesium oxide glass at various wavelengths to find the minimum transmittance point and adjust if necessary.