Hengping 721/721N Visible Spectrophotometer User Manual

This document provides detailed information about the principles, uses, and characteristics of the 721/721N visible spectrophotometer. It also includes specifications and standard accessories of the instrument. The document provides a guide for operating the instrument, including installation conditions and unboxing inspection. Additionally, the document covers basic operation and preheating process of the instrument.


Hengping 721/721N Visible Spectrophotometer User Manual - cover page
Brand
Hengping
Document type
User Manual
Language
Chinese
Pages
14
File format
PDF
File size
169 KB
Published
09 December, 2022
Updated
27 September, 2023
MD5 checksum
92298122192C00527A0B4E19AB528A86

About this manual

The document serves as a user guide for the Hengping 721/721N Visible Spectrophotometer. It details the instrument's specifications, including its wavelength range of 335 to 1000nm and optical system components like the holographic grating and halogen lamp source.

The manual covers various operational modes such as transmittance, absorbance, concentration factor, and direct reading concentration. It provides step-by-step instructions for setting up the instrument, performing zero and 100% adjustments, calibrating wavelength accuracy using magnesium oxide glass, and troubleshooting common faults. This manual is written in Chinese.

What's inside

  • Instrument Specifications
  • Operating Modes
  • Key Functions
  • Installation and Setup
  • Wavelength Calibration
  • Transmittance Adjustment
  • Concentration Factor Setting
  • Troubleshooting

Specifications

Wavelength Range335 ~ 1000nm
Optical SystemC-T type monochromator, 1200 lines/mm holographic grating
Light SourceHalogen lamp (COW/1 12V)
Wavelength Accuracy+/- 2nm
Photometric Accuracy+/- 1.0%(T)
Power Supply220V+10% 50Hz

Frequently asked questions

What is the wavelength range of this spectrophotometer?

The instrument operates within a wavelength range of 335 to 1000nm.

How long should the instrument be preheated before use?

The manual recommends a preheat time of not less than 30 minutes to reach thermal equilibrium.

What is the procedure for calibrating wavelength accuracy?

To calibrate, set the display to 800nm with a blank sample, then measure magnesium oxide glass at various wavelengths to find the minimum transmittance point and adjust if necessary.