The PAT-80 ● PAT-85 ● PAT-86 tester is a modern, high-quality tester, easy and safe in operation. Please acquaint yourself with the manual in order to avoid measuring errors and prevent possible problems related to operation of the tester.
The SONEL PAT-80, PAT-85, and PAT-86 are modern, high-quality digital testers designed to check the safety status of electrical equipment. The device measures basic parameters such as protective conductor resistance, insulation resistance, continuity, leakage current, and RCD effectiveness in accordance with various international standards including EN 60745-1, EN 60335-1, and VDE 0404.
| Input voltage range | 195...265 V, 50 Hz |
|---|---|
| Protective fuse rating | Two 16 A fuses |
| Insulation resistance range | up to 599 MΩ |
| Leakage current upper limit (I_SUB, I_PE) | 0.01 mA ... 9.9 mA |
| Touch leakage current upper limit | 0.01 mA ... 1.99 mA |
For mains voltages below 195 V and above 256 V all test functions are automatically blocked.
No, it must only be operated by appropriately qualified personnel to avoid damage or danger.
Do not start measurements until the tester has warmed up to ambient temperature for approximately 30 minutes.