Thermo Electron NZP1000 Series Nozzle Pitot Air/Gas Flow Sensor Data Sheet

The NZP1000 Series Nozzle Pitot Air/Gas Flow Sensor features no upstream straight run, ±0.5% reading accuracy, excellent repeatability, and a 34:1 flow turndown ratio.


Thermo Electron NZP1000 Series Nozzle Pitot Air/Gas Flow Sensor Data Sheet - cover page
Brand
Thermo Electron
Category
Other
Document type
Product Leaflet
Language
English
Pages
6
File format
PDF
File size
741 KB
Updated
28 September, 2023
MD5 checksum
A8776CAA6DFD4ECF2056A1DB172AE5E3

About this manual

The Thermo Electron NZP1000 Series Nozzle Pitot Air/Gas Flow Sensor is a high-accuracy flowmeter designed specifically for circular ducts. It combines integral flow straighteners, a nozzle, and a pitot averaging array to measure air and gas flow without requiring an upstream straight run. This design allows it to perform accurately even in conditions with irregular flow profiles and turbulence where conventional primaries might fail.

The document provides detailed performance specifications, including ±0.5% reading accuracy, ±0.1% repeatability, and a 34:1 flow turndown. It outlines various configuration options for materials (Carbon Steel, Stainless Steel), flanges, and accessories such as temperature and pressure connections. Installation guidelines and physical dimensions for different nominal duct sizes are also included.

What's inside

  • Performance Specifications
  • Functional Specifications
  • Physical Specifications
  • Options
  • Installation Guidelines
  • Specifying the Brandt NZP1000
  • Proven Applications
  • Dimensions

Specifications

AccuracyTypically ±0.5% of Flow
Repeatability±0.1% of Flow
Velocity Turn-down34:1
Recommended Velocity Range175 - 6000 FPM
Temperature Rating (Carbon Steel)-200° to 750°F
Maximum Static Pressure (Low Pressure Carbon Steel)10 psig

Frequently asked questions

Does the NZP1000 Series require an upstream straight run?

No, it requires no upstream straight run and only 1-2 downstream diameters to maintain accuracy.

What is the permanent pressure loss of the sensor?

The permanent pressure loss is approximately 44% of the measured differential pressure.

What materials are available for the flow straighteners?

Options include Aluminum, 304 Stainless Steel, 316 Stainless Steel, or special materials upon consultation.

Download PDF (741 KB)