Agilent 1100 Series Amino Acid Analysis Technical Note

This technical note demonstrates the performance of amino acids analysis on the Agilent 1100 Series modules and systems for LC. Detailed information regarding instrumental and chromatographic conditions and performance are given. The precision for retention times over 10 runs is as low as 0.2 % RSD, and the RSD of areas is between 0.6 and 5 %. Since the introduction of automated amino acid analysis on the HP 1090 Series HPLC system using a two step precolumn derivatization, this technique has become a well accepted and routine method for the analysis of primary and secondary amino acids.


Agilent 1100 Series Amino Acid Analysis Technical Note - cover page
Brand
Sensitive
Category
Other
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
795 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
32AE043183CB147E5E66D04A8D43C72F

About this manual

This document details the performance of amino acid analysis on the Agilent 1100 Series modules and systems for Liquid Chromatography (LC). It demonstrates that this technique can achieve better performance than the previous HP 1090 Series system. The note covers instrumental and chromatographic conditions, including the use of binary pumps, online vacuum degassers, autosamplers, thermostatted column compartments, and detectors (DAD and FLD).

The document provides specific experimental procedures for preparing mobile phases, standards, and derivatization reagents using ortho-phthalaldehyde (OPA) and 9-fluorenylmethyl chloroformate (FMOC). It includes detailed instructions for preparing calibration standards for both standard and high sensitivity analyses, as well as extended amino acid kits. Results are presented for both UV and fluorescence detection methods, highlighting the precision of retention times and areas, and the limits of detection.

What's inside

  • Introduction
  • HPLC Instrumentation
  • Amino Acid Standards
  • Derivatization Reagents
  • Preparing Mobile Phases
  • Preparing Internal Standards
  • Preparing Calibration Standards
  • Extended Amino Acids
  • Results
  • Chromatographic Conditions

Specifications

Limit of Detection (DAD)below 5 pmol
Limit of Detection (FLD)below 100 fmol
Linearity Correlation Factor0.99998 to 0.99999
Retention Time Precision0.2 % RSD
Area Precision (RSD)0.6 to 5 %
Flow Rate0.45 ml/min
Oven Temperature40 C

Frequently asked questions

Which detector is recommended for concentrations below 100 pmol?

The fluorescence detector (FLD) is recommended for concentrations below 100 pmol due to better selectivity.

What reagents are used for derivatization?

Ortho-phthalaldehyde (OPA) is used for primary amino acids and 9-fluorenylmethyl chloroformate (FMOC) is used for secondary amino acids.

What is the flow rate for the chromatographic conditions?

The flow rate is set to 0.45 ml/min.