Hewlett-Packard G1005A N-terminal Routine 3.1/3.5 Sequencer Methods Technical Note

This document describes the implementation of the Edman degradation chemistry on Hewlett-Packard's column-based protein sequencer and its applicability for N-terminal amino acid sequence analysis of most protein and peptide samples in the nanomole to low picomole range.


Hewlett-Packard G1005A N-terminal Routine 3.1/3.5 Sequencer Methods Technical Note - cover page
Brand
N-Terminal
Document type
Service Manual
Language
English
Pages
8
File format
PDF
File size
51 KB
Published
05 July, 2012
Updated
28 September, 2023
MD5 checksum
0010B490DB9F2C80310F76C20DAC4B75

About this manual

This document details the Routine 3.1/3.5 sequencer methods used with Hewlett-Packard's column-based protein sequencer, specifically model G1005A. These methods implement Edman degradation chemistry for N-terminal amino acid sequence analysis of protein and peptide samples in the nanomole to low picomole range.

What's inside

  • Routine column method
  • Routine converter method
  • Routine Cycle 1 (column)
  • Routine PTH Std (converter)
  • Column Prep method
  • Reagent bottle configuration
  • Routine 3.1 column method steps
  • Routine 3.5 column method steps

Specifications

Sample rangenanomole to low picomole range
Routine 3.1 column method time43.9 min (2631.6 sec)
Routine Cycle 1 time58.7 min (3520.1 sec)
Routine 3.5 column method time42.0 min (2518.0 sec)
Waste volume per cycleless than 1mL/cycle

Frequently asked questions

What reagents are required for the Routine 3.1 method?

The G2046A kit is purchased from Hewlett-Packard for the Routine 3.1 method.

What reagents are required for the Routine 3.5 method?

The G2021A kit is purchased from Hewlett-Packard for the Routine 3.5 method.

What is the purpose of the Column Prep method?

It is used to wash the sequencer column prior to sample loading, which is especially important for low-level samples.